Linear accelerator workstation software suite

This recall has been terminated (originally issued July 20, 2004).

Company
Varian Medical Systems, Inc.
Recall Initiated
May 27, 2004
Posted
July 20, 2004
Terminated
December 20, 2004
Recall Number
Z-1084-04
Quantity
155 units
Firm Location
Palo Alto, CA
Official Source
View on FDA website ↗

Reason for Recall

An anomaly occurs in the software suite (Millennium MLC Workstation v 6.3 and 6.4, Integrated Treat v.6.5) when used in combination with the standard series MLC controller software v 5.0 or 5.1

Distribution

units were distributed throughout the US and Canada, as well as approximately 66 other foreign consignees, including

Lot / Code Info

MLC Workstation v 6.3 and 6.4, Integrated Treat v.6.5) when used in combination with the standard series MLC controller software v 5.0 or 5.1

Root Cause

Software design

Action Taken

Varian will notify customers via phone beginning 5/27/2004 and distribute a product notification letters June 3, 2004, notifying them of the anomaly and corrective action, and finally, the anomaly will be corrected by a future release of software.