RecallDepth

ACCU-CHECK PERFORMA STRIP NO. 05987288031, 100 BANDELETTES REACTIVES, ROCHE DIAGNOSTICS GMBH, WWW.ACCUCHEK.COM glucose monitoring test strips.

This recall has been terminated (originally issued June 21, 2011).

Company
Roche Diagnostics Operations, Inc.
Recall Initiated
May 18, 2011
Posted
June 21, 2011
Terminated
October 25, 2011
Recall Number
Z-2570-2011
Quantity
27,019 units
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

Erroneous results: Internal investigations revealed that the test strips exhibited a performance issue that can cause test strip errors or a bias in test result that exceed the tolerance range according to EN ISO 15197.

Distribution

No US distribution, in France Only.

Lot / Code Info

Lot Number 470049, expiration date 12/31/2011.

Root Cause

Component design/selection

Action Taken

Direct accounts were notified by a Roche Diagnostics press release on 5/18/2011 that included the product's name, lot number, and information to withdraw the remaining inventory of lot 470049 from storage