Comprehensive Shoulder Instruments Fracture Stem Inserter w/Version Control, REF 31-406901, Biomet Orthopedics Warsaw, IN, Instrument used to introduce the stem into the canal. The instrument is used to insert the shoulder implant into the shoulder.

This recall has been terminated (originally issued June 30, 2011).

Company
Biomet, Inc.
Recall Initiated
April 7, 2011
Posted
June 30, 2011
Terminated
April 8, 2013
Recall Number
Z-2713-2011
Quantity
469 both products
Firm Location
Warsaw, IN
Official Source
View on FDA website ↗

Reason for Recall

When the surgeon attempts to remove the inserter from the stem, the inserter may become stuck not allowing it to disengage from the stem. The stem may have to be removed from the patient in order to disengage the inserter from the stem. If the stem is removed from the patient, it could potentially cause a significant delay in the procedure or damage/fracture of the humerus which may require inter

Distribution

Worldwide distribution: USA (nationwide) including Puerto Rico; and countries of: Australia, Canada, Chile, Columbia, Costa Rica, Italy, Japan, Korea, Netherlands, New Zealand, Spain and Sweden.

Lot / Code Info

064320, 189990, 293840, 299010, 299020, 401800, 401810, 406610, 444560, 444600, 517880, 517960, 525360, 534260, 657870, 723150 and  727030.

Root Cause

Nonconforming Material/Component

Action Taken

The firm, Biomet Orthopedics, sent two "URGENT MEDICAL DEVICE RECALL NOTICE" letters one dated April 7, 2011 and the second letter dated April 12, 2011 which included an additional lot number, to its customers. The letters described the product, problem and actions to be taken. The customers were instructed to immediately locate and discontinue use of the noted product; carefully follow the instructions on the enclosed "FAX Back Response Form"; complete and return the Fax Response Form within three (3) business days via fax to 574-372-1683; use priority carrier for their shipment; If the product was further distributed, they MUST notify and give letter to Hospital personnel, and confirm receipt of this notice by calling, 800-348-9500, extension 3755 or 3756. Questions related to this notice should be directed to 574-371-3755 or 574-371-3756, Monday through Friday, 8AM to 5PM.