RecallDepth

Hamilton-G5 Ventilator with software version V2.00x and Neonatal Option or V2.01x and Neonatal software. Hamilton-G5: catalog number 159001; Neonatal Software catalog number: 159700; Neonatal Option catalog number: 159187. Hamilton Medical Intended for use in the hospital and institutional environment where healthcare professionals provide patient care.

Company
Hamilton Medical Inc
Recall Initiated
February 16, 2011
Posted
April 8, 2011
Terminated
March 21, 2012
Recall Number
Z-1928-2011
Quantity
97 units
Firm Location
Reno, NV

Reason for Recall

The adaptive volume/pressure regulation can be automatically and permanently reduced to PEEP + 3 mbar when a closed suctioning maneuver is performed on the patient and may be undetected by the operator.

Distribution

Nationwide Distribution -- MA, FL, GA, NC, MI, NC, MO, TX, and WV.

Lot / Code Info

Software version V2.00x and Neonatal Option or V2.01x and Neonatal software

Root Cause

Software design

Action Taken

A Medical device Safety Alert and Corrective Action along with a cover letter, was sent to all locations on February 16, 2011. The letter identified the affected product along with the reason for recall. The letter states that operators deactivate the "high tidal volume" alarm as described in appendix K.4 of the operator's manual until the ventilator is upgraded with a corrected software. Distributors will upgrade the affected devices with the corrected software as soon as possible. Questions can be directed to 800-426-6331, ext 215.

More recalls by Hamilton Medical Inc

View all recalls by this company →