IMx Tacrolimus II Assay, list 3C10-20; Abbott Diagnostics, a Division of Abbott Laboratories, Abbott Park, IL 60064

Company
Abbott Laboratories Hpd/Add/Gprd
Recall Initiated
November 13, 2003
Posted
December 17, 2003
Terminated
October 29, 2004
Recall Number
Z-0230-04
Quantity
not lot specific
Firm Location
Abbott Park, IL

Reason for Recall

Inadequate instructions for use resulting in calibration errors and the protein interference data in the package insert is not representative of assay performances.

Distribution

Nationwide and internationally to Guatemala, Venezuela, Columbia, Peru, Chile, Australia, Panama, Argentina, Puerto Rico, Dominican Republic, Canada, Germany, New Zealand, England, Japan, Singapore, Thailand, Korea and Costa Rica.

Lot / Code Info

All lot numbers

Root Cause

Other

Action Taken

Device correction recall initiated by letters dated 11/12/03. Letters were sent to all current IMx Tacrolimus II customers worldwide, advising them that calibration errors and controls out of range be the result of variance with the sample pretreatment steps listed in the package insert, and that the protein interference data in the insert is not representative of assay performance (12% when it should state less than 22% error) . A Customer Training Guide was included in the letter, outlining steps to aid the assay user in performing the pretreatment step and listing the most common deviations that can impact results.