Hamilton brand Disposable Precision Tips, Part Number 235300 (box containing 504 tips) and 235400 (case containing 20 boxes), Manufactured by Hamilton Company, Reno, Nevada
This recall has been terminated (originally issued April 7, 2007).
- Company
- Hamilton Co
- Recall Initiated
- January 30, 2007
- Posted
- April 7, 2007
- Terminated
- November 28, 2007
- Recall Number
- Z-0723-2007
- Quantity
- 37 cases
- Firm Location
- Reno, NV
- Official Source
- View on FDA website ↗
Reason for Recall
Some of the packaged tips have increased force during aspiration, which can lead to pipetting errors such as LLD error messages, and also has risk of causing a missed pipetting.
Distribution
Nationwide (via a single distributor in New Jersey)
Lot / Code Info
Lot: 33906 P5 and Lot: 63906 P5
Root Cause
Other
Action Taken
The firm initiated its recall on January 30, 2007. The initial notification was sent via certified mail and e-mail. Subrecalls will be conducted by the distributor.