Finish Thompson Centrifugal Pump Model #110192-3 (sold as spare part)

Company
Isopure Corp
Recall Initiated
April 19, 2019
Posted
January 19, 2020
Terminated
May 29, 2020
Recall Number
Z-0823-2020
Quantity
8
Firm Location
Louisville, KY

Reason for Recall

There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.

Distribution

The products were distributed to the following US states: AL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, and VA.

Lot / Code Info

Pump Serial Numbers: 178323B18, 178751C18, 178749C18, 180214D18, 180213D18, 180208D18, 178747C18, 178750C18

Root Cause

Nonconforming Material/Component

Action Taken

The only affected customer was contacted via telephone on 4/19/19. An URGENT PRODUCT FIELD CORRECTION NOTICE dated 5/22/19 was distributed to affected clinics.