Medline medical procedure convenience kits labeled as: 1) INPATIENT CTRL LN DRG CHG, Kit SKU DT22880; 2) ADULT CVC DRSG CHG W/ADHESIVE, Kit SKU DYNDC3283A.
- Recall Initiated
- December 12, 2025
- Posted
- January 28, 2026
- Recall Number
- Z-1195-2026
- Quantity
- 1,928 total
- Firm Location
- Northfield, IL
Reason for Recall
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Distribution
US Nationwide distribution.
Lot / Code Info
Medline Kit SKU DT22880: UDI/DI 10653160995470 (EA) 40653160995471 (CS), Lot Number 25KMJ441; Medline Kit SKU DYNDC3283A: UDI/DI 10198459313240 (EA) 40198459313241 (CS), Lot Number 25KMH515.
Root Cause
Nonconforming Material/Component
Action Taken
Medline Industries issued a URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/12/2025 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-25-251-FGX2 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product." For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.