Abre Venous Self-Expanding Stent System, Model numbers AB9G14120090 AB9G12100090 AB9G14080090 AB9G12080090 AB9U12060090 AB9U14100090 AB9U14080090 AB9U12080090 AB9U14060090 AB9G14100090 AB9G12060090

Company
Medtronic Inc.
Recall Initiated
December 4, 2025
Posted
January 15, 2026
Recall Number
Z-1098-2026
Quantity
442
Firm Location
Plymouth, MN

Reason for Recall

Stents may be impacted by a Nitinol anomaly which causes a reduced stent diameter when deployed and corresponding reduced radial force. Issue may result in stent migration.

Distribution

Worldwide - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Finland, Germany, Jordan, Lebanon, Malaysia, Netherlands, New Zealand, Peru, Poland, Russia, Singapore, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Uzbekistan, Vietnam.

Lot / Code Info

UDI-DI (GTIN): 00763000547332 00763000547271 00763000547318 00763000547264 00643169796225 00643169796300 00643169796294 00643169796232 00643169796287 00763000547325 00643169796300 00763000547257 Serial/Lot numbers C112511 C113253 C113795 C117416 C117482 C121084 C123953 C133327 C133329 C134478 C138060 C150325

Root Cause

Under Investigation by firm

Action Taken

Beginning 04-Dec-2025, Medtronic distributed an Urgent Medical Device Recall notice to all U.S. consignees who have impacted products via mail courier. Consignees were instructed to complete and return a response form to the firm and to return any affected product they may have.