RecallDepth

Medline RenewalLigasure Blunt Tip Sealer/Divider, NanoCoated Compatible with FT10 Generator 23 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed

This recall is currently active, issued November 7, 2025.

Recall Initiated
October 1, 2025
Posted
November 7, 2025
Recall Number
Z-0472-2026
Quantity
65 units
Firm Location
Northfield, IL
Official Source
View on FDA website ↗

Reason for Recall

Medline Renewal has determined that certain areas on the device have the potential to trap residual material, which may remain within the device and, in rare cases, become dislodged.

Distribution

US Nationwide distribution.

Lot / Code Info

UDI/DI 10888277680371 (each), 20888277680378 (case), All Lots

Root Cause

Under Investigation by firm

Action Taken

Medline issued an URGENT MEDICAL DEVICE RECALL notice to its consignees via email and USPS first class mail. The notice explained the issue, potential risk to health, and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-25-206 Recall Code: 3. Your account will receive credit once the response form is submitted. 4. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com."