C-Rex UNO Microinterventional Goniotomy Instrument. Model Number: CRX-120. Product Description: The C-Rex UNO Microinterventional Goniotomy Instrument is a sterile single use disposable manual surgical instrument used in ophthalmic procedures for ab interno goniotomy and inner wall trabeculotomy.

This recall is currently active, issued October 10, 2025. It was issued by Iantrek Inc..

Company
Iantrek Inc.
Recall Initiated
September 6, 2025
Posted
October 10, 2025
Recall Number
Z-0135-2026
Quantity
198
Firm Location
Providence, RI
Official Source
View on FDA website ↗

Reason for Recall

Potential for compromised integrity of the sterile packaging.

Distribution

US Nationwide distribution in the states of AL, AZ, CA, CT, DE, FL, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI.

Lot / Code Info

1. Model Number: CRX-120; UDI: +B962CRX1200/$$7351729/16D20250501/14D20260501/Q1$; Lot Number 351729; Exp date 05/2026. 2. Model Number: CRX-120 UDI: +B962CRX1200/$$7351759/16D20250501/14D20260501/Q1%; Lot Number: 351759; Exp date 05/2026.

Root Cause

Process control

Action Taken

Iantrek notified consignees on 09/06/2025 via email. Consignees were instructed to immediately remove and quarantine all affected units to prevent their use, notify personnel or customers if affected units were forwarded to other facilities. Consignees may also alternatively provide Iantrek with contact information so that they can notify them on the customer's behalf. An Iantrek representative will be contacting each consignee immediately to help facilitate removal, return and replacement of affected units. They will arrange to have replacement product shipped to consignees at no cost.