PERIFIX FX Continuous Epidural Anesthesia Tray with PERIFIX 17 Ga. x 3-1/2 in. (8.9cm) Tuohy Epidural Needle - Winged and 19 Ga. Closed tip PERIFIX FX Springwound Epidural Catheter. Product Code: CE17TKFCPS. Product Description: CE17TKFCPS EPIDURAL FULL TRAY.

Company
B Braun Medical Inc
Recall Initiated
September 18, 2025
Posted
October 9, 2025
Recall Number
Z-0115-2026
Quantity
0 units
Firm Location
Allentown, PA

Reason for Recall

Potential for the lid of the catheter connector to be in the incorrect position.

Distribution

Nationwide distribution to AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV.

Lot / Code Info

REF: 332078; Product Code: CE17TKFCPS; UDI-DI (Primary): 04046955899432; UDI-DI (Unit of Use): 04046955899449; Lot Numbers: 0062006314, 0062011670, 0062028428. Expiration Date: 07/31/2026

Root Cause

Process control

Action Taken

B Braun notified consignees on about 09/18/2025 via USPS certified mailed URGENT MEDICAL DEVICE CORRECTION NOTIFICATION. Consignees were instructed to, where possible, have a second kit or additional catheter connector available at the point of use and upon opening the kit, verify the correct catheter connector is present. If the affected component is detected, discard and do not use. Consignees were also instructed to review the Medical Device Correction Notification and ensure all users in your organization and other concerned persons are informed, if affected units have been further distributed, notify customers, post the notification where affected units are stored, and complete and return the "Urgent Medical Device Correction Acknowledgement Form." Additionally, if consignees have identified affected units in inventory, they may order one case (50 pieces) of affected sterilized single use catheter connectors to have on hand.