iBOT PMD with software version 01.05.24. Personal Mobility Device.

This recall is currently active, issued June 12, 2025. It was issued by Mobius Mobility Llc.

Company
Mobius Mobility Llc
Recall Initiated
April 24, 2025
Posted
June 12, 2025
Recall Number
Z-1965-2025
Quantity
442 units
Firm Location
Manchester, NH
Official Source
View on FDA website ↗

Reason for Recall

Software issue that could potentially lead to the device tipping over from Balance Mode.

Distribution

US Nationwide distribution in the states of CA, CO, FL, GA, IN, NC, NJ, OH, PR, RI, TN, TX & VA.

Lot / Code Info

Software Version 01.05.24; UDI (01) 00857584008010 (21) 0001108, (01) 00857584008010 (21) 0001110, (01) 00857584008010 (21) 0001121, (01) 00857584008010 (21) 0001123, (01) 00857584008010 (21) 0001128, (01) 00857584008010 (21) 0001136, (01) 00857584008010 (21) 0001137, (01) 00857584008010 (21) 0001422, (01) 00857584008010 (21) 0001438, (01) 00857584008010 (21) 0001496, (01) 00857584008010 (21) 0001505, (01) 00857584008010 (21) 0001508, (01) 00857584008010 (21) 0001516, (01) 00857584008010 (21) 0001517, (01) 00857584008010 (21) 0001519; Serial No. 0001108, 0001422, 0001438, 0001496, 0001505, 0001508, 0001516, 0001517, 0001519, 0001110, 0001121, 0001123, 0001128, 0001136, 0001137.

Root Cause

Software design

Action Taken

Consignees were initially contacted via telephone on 4/24/25 to notify them of this software issue. An URGENT MEDICAL DEVICE CORRECTION NOTICE dated 5/5/25 was emailed to consignees to notify them of the recall and instructing them that the firm is developing a software correction. Once Mobius Mobility has obtained a correction they will reach out to consignees to schedule a time to update affected devices. Consignees with any questions can email service @mobiusmobility.com or call 833-346-4268x2.