RecallDepth

Brand Name: HALYARD Product Name: Laparotomy Pack Model/Catalog Number: 88231 Software Version: N/A Product Description: Laparotomy Pack Component: N/A

This recall is currently active, issued January 17, 2025. It was issued by O & M Halyard, Inc..

Company
O & M Halyard, Inc.
Recall Initiated
November 26, 2024
Posted
January 17, 2025
Recall Number
Z-0983-2025
Quantity
1440 units
Firm Location
Mechanicsville, VA
Official Source
View on FDA website ↗

Reason for Recall

Elevated residues of anti-block/slip additive used in manufacturing found within sterile packaging.

Distribution

US Nationwide distribution in the states of CO, FL, GA, HI, IA, IL, IN, KS, LA, MI, MN, MO, NC, NE, NY, TN, TX, VA, WI.

Lot / Code Info

Model No 88231; UDI-DI: 30680651882319; Lot Number: AC2335502D; Exp Date 21-12-28

Root Cause

Process control

Action Taken

On November 26, 2024 URGENT VOLUNTARY RECALL letters were emailed to consignees. ACTIONS TO TAKE: " Immediately discontinue use of the Laparotomy Pack identified in APPENDIX 1. Please review your inventory, and segregate and quarantine all affected Packs. If any of the affected trays identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. " Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. " Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. " RETURN the completed Field Action Response Form to GM-OMRA_RECALLS@owens-minor.com as soon as possible even if you have no current product inventory. ADDITIONAL INFORMATION: Report any quality problems or adverse events associated with the products listed in this notification to complaints@owens-minor.com. Your assistance is appreciated and necessary to prevent any risk to patient health as a result of not following the field correction instructions. If you have any questions, please email GM-OMRA_RECALLS@owens-minor.com using the Event # FA-2024-031 in the header of the response. This will ensure prompt response by the O&M Halyard, Inc, team.