RecallDepth

Halyard CPT PEDS ADULT CH PACK - Medical convenience kits Model Number: SHAN431-06

Company
Avid Medical, Inc.
Recall Initiated
September 20, 2024
Posted
November 14, 2024
Recall Number
Z-0432-2025
Quantity
81 eaches
Firm Location
Toano, VA

Reason for Recall

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Distribution

US Nationwide distribution.

Lot / Code Info

GTIN: 10809160417126 Lot Number: 1599174

Root Cause

Process control

Action Taken

Avid Medical issued Urgent: Voluntary Medical Device Recall letter via email on 9/20/24. Letter states reason for recall, health risk and action to take: If you received any of the batches/lots identified in APPENDIX 1, please review your inventory, and discard the entire kit. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. Distributors: If any of the affected kits identified in Appendix 1 were further distributed, please ensure all end-users are appropriately notified of this product correction and maintain records of effectiveness. "Complete the attached Recall Response Form and return as soon as possible. Report any quality problems or adverse events associated with the products listed in this notification to complaints@owens-minor.com.

More recalls by Avid Medical, Inc.