2090 Programmers with serial number prefixes PKK0 and PKK1: a) 2090, Product Description: PRGM 2090 PROGRAMMER-PORTABLE CRM; b) 209031, Product Description: PROGRAMMER 209031 PORTABLE CRM; c) 2090AA, Product Description: PROGRAMMER 2090AA PORTABLE CRM; d) 2090AA4, Product Description: 2090AA4 PROGRAMMER-PORTABLE CRM; e) 2090AC, Product Description: SFTWR 2090AC PROGRAMMER-PORTABLE CRM W/; f) 2090AM, Product Description: PROGRAMMER CARELINK HONG KONG ASIA; g) 2090CHEN, Product Description: PRGM 2090CHEN PKG-ASSY PROGRAMMER 2090; h) 2090CHFR, Product Description: PRGM 2090CHFR PKG-ASSY PROGRAMMER 2090; i) 2090CL, Product Description: PROGRAMMER 2090CL INTL-JAPAN MKT; j) 2090EN, Product Description: PRGM 2090EN PKG-ASSY PROGRAMMER 2090 UK; k) 2090FR, Product Description: PRGM 2090FR PKG-ASSY PROGRAMMER 2090 IN; l) 2090LDEN, Product Description: PRGM 2090LDEN PROGRAMMER-PORTABLE LOANE; m) 2090NL, Product Description: PRGM 2090NL PKG-ASSY PROGRAMMER 2090 IN; n) 2090W, Product Description: PROGRAMMER 2090W TELEMETRY C; o) 2090WCHDE, Product Description: PROGRAMMER 2090WCHDE SWISS-GERMAN; p) 2090WCHEN, Product Description: PROGRAMMER 2090WCHEN SWISS-ENGLISH; q) 2090WCHFR, Product Description: PROGRAMMER 2090WCHFR SWISS-FRENCH; r) 2090WDE, Product Description: PROGRAMMER 2090WDE GERMAN MKT; s) 2090WFR, Product Description: PROGRAMMER 2090WFR FRENCH MKT; t) 2090WLD, Product Description: PROG 2090WLD ALL LANGUAGES W/ANALYZER; u) 2090WNL, Product Description: PROGRAMMER 2090WNL NETHERLANDS MKT
This recall is currently active, issued September 13, 2024. It was issued by Medtronic Inc..
- Company
- Medtronic Inc.
- Recall Initiated
- July 11, 2024
- Posted
- September 13, 2024
- Recall Number
- Z-3152-2024
- Quantity
- 18788 units
- Firm Location
- Mounds View, MN
- Official Source
- View on FDA website ↗
Reason for Recall
Medtronic CareLink 2090 programmers with serial number prefixes PKK0 and PKK1 have the potential for autonomous cursor motion when Finger Touchscreen capability is enabled by software version 3.2 or higher.
Distribution
Worldwide distribution.
Lot / Code Info
2090 Programmers with serial number prefixes PKK0 and PKK1: a) GTIN: 00613994136480, 00613994136497, 00613994136503, 00613994136510, 00613994136527, 00613994136534, 00613994136565, 00613994136596, 00613994136619, 00613994136626, 00613994175076, 00613994175083, 00613994175090, 00613994175106, 00613994175120, 00613994175137, 00613994175144, 00613994175151, 00613994175168, 00613994175175, 00613994175182, 00613994175205, 00613994175212, 00613994175229, 00613994175236, 00613994175243, 00613994175250, 00613994175267, 00613994175281, 00613994175298, 00613994175304, 00613994175311, 00613994175328, 00613994175335, 00613994175342, 00613994175366, 00613994175373, 00613994175380, 00613994175397, 00613994175403, 00613994175410, 00613994175434, 00613994175465, 00613994175489, 00613994175540, 00613994175588, 00613994175595, 00613994203182, 00613994360922, 00613994434883, 00613994434890, 00613994434944, 00613994435002, 00613994435255, 00613994435316, 00613994580863, 00613994580894, 00613994580931, 00613994580962, 00613994581013, 00613994581051, 00613994581075, 00613994581105, 00613994581150, 00613994581167, 00613994581174, 00613994581181, 00613994581204, 00613994581211, 00613994581228, 00613994581235, 00613994581242, 00613994581549, 00613994581570, 00613994581587, 00613994581594, 00613994581624, 00643169440128, 00643169440166, 00643169440173, 00643169440197, 00643169440203, 00643169440937, 00643169537750, 00643169537774, 00643169537811, 00643169537828, 00643169537835, 00643169537842, 00643169537859, 00643169537866, 00643169537880, 00643169537910, 00643169537927, 00643169538016, 00643169629004, 00643169629011, 00643169629028, 00643169629035, 00643169629042, 00643169629066, 00721902246227, 00721902246234, 00721902246241, 00721902246265, 00721902246289, 00721902246302, 00721902246326, 00721902246333; b) GTIN: 00721902171222; c) GTIN: 00721902171079; d) GTIN: 00721902824692; e) GTIN: 00721902059100; f) GTIN: 00721902171154; g) GTIN: 00721902246357; h) GTIN: 00721902246364; i) GTIN: 00721902480201; j) GTIN: 00721902246159, 00721902246173; k) GTIN: 00721902246203; l) GTIN: 00721902841903; m) GTIN: 00721902246272; n) GTIN: 00613994127631, 00613994435354, 00613994581549, 00885074301440, 00885074310596, 00885074310633, 00885074310657, 00885074310930, 00885074312866, 00885074312873, 00885074313047, 00885074313108, 00885074313115, 00885074313139, 00885074313146, 00885074313153, 00885074313160; o) GTIN: 00885074313191; p) GTIN: 00885074313177; q) GTIN: 00885074313184; r) GTIN: 00885074314921; s) GTIN: 00885074313009; t) GTIN: 00885074357416; u) GTIN: 00885074313122
Root Cause
Device Design
Action Taken
Medtronic issued a MEDICAL DEVICE CORRECTION notice to its consignees on 07/11/2024 via email or UPS 2-day delivery. The notice explained the issue, potential risk, and provided the following: "Instrument Management Recommendations Software updates are necessary to maintain proper programmer function. Medtronic representatives will assist in performing the software update on all Medtronic CareLink 2090 programmers and assess proper function. Medtronic representatives will assist with returning programmers needing repair or replacement. " Following review of this letter, please complete the enclosed Confirmation Form and email it to RS.FCACRM@medtronic.com to confirm receipt of this notification. " Please forward this notice to others who need to be aware within your organization."