Endoskeleton" TAS Interbody System with Titan nanoLOCK" Surface Technology, Nanolock 12 Dg Interbody, Large, 12mm

Company
Medtronic Sofamor Danek Usa Inc
Recall Initiated
September 2, 2021
Terminated
October 5, 2023
Recall Number
Z-0085-2022
Quantity
10 units
Firm Location
Memphis, TN

Reason for Recall

Engraving on implants may not match what is listed on the pouch labels.

Distribution

Distribution in US: CA, MN, MO, NC, NV, NY, OH, OK, OR, PA, SC, SD, WA, WV

Lot / Code Info

Product Number 2312-0212-N, Lot A200323 GTIN 00191375014171

Root Cause

Packaging change control

Action Taken

A customer letter was sent via FedEx to impacted US consignees and delivered on or about September 3, 2021. Customers are asked to locate and remove the impacted product from normal storage locations. Customers are advised to not use product. A Customer Confirmation Form should be completed and returned to rs.sdmriskmgt@medtronic.com. Customers may also coordinate return and replacement of impacted product with the Medtronic Sales Representative. This recall notice should be forwarded to those within the customer organization or to any organization where potentially impacted devices have been transferred.