Semi-automatic Biopsy-Needle, ITP innotom.com, CE 0297, STERILE EO for the following article numbers: Article no. BIM 18/10, Length 100 mm, Diameter 18 G (1.25 mm); Article no. BIM 18/15, Length 150 mm, Diameter 18 G (1.25 mm); Article no. BIM 18/20, Length 200 mm, Diameter 18 G (1.25 mm); Article no. BIM 16/10, Length 100 mm, Diameter 16 G (1.6 mm); Article no. BIM 16/15, Length 150 mm, Diameter 16 G (1.6 mm); Article no. BIM 16/20, Length 200 mm, Diameter 16 G (1.6 mm); Article no. BIM 14/10, Length 100 mm, Diameter 14 G (2.1 mm); Article no. BIM 14/15, Length 150 mm, Diameter 14 G (2.1 mm); Article no. BIM 14/20, Length 200 mm, Diameter 14 G (2.1 mm) *NOT DISTRIBUTED WITHIN THE US

Company
Innovative Tomography Product Gmbh Universitatsstr. 136 Bochum Germany
Recall Initiated
April 21, 2021
Recall Number
Z-2164-2021

Reason for Recall

Sterility Assurance; Coaxial Interventional Needle labeled as sterile may not have been adequately sterilized

Distribution

US Nationwide distribution in the state of CA.

Lot / Code Info

Affected Lot Numbers:  BIM 18/10: LOT: 4918;  BIM 18/15: LOT: 4017, 4918, 4719;  BIM 18/20: LOT: 4017, 4918;  BIM 16/10: LOT: 4918, 1119, 1219;  BIM 16/15: LOT: 4918, 1119;  BIM 16/20: LOT: 4918, 1119;  BIM 14/10: LOT: 4918, 3019;  BIM 14/15: LOT: 4918;  BIM 14/20: LOT: 4918

Root Cause

Vendor change control

Action Taken

On April 20, 2021, Innovative Tomography Products GmbH issued a "urgent Safety Information Product Recall" notification to affected consignees via E-Mail. In addition to informing consignees about the recall products, the firm asked consignees mo to take the following actions: 1. We request that you immediately check your stock of the products in question with the LOT numbers indicated. Please block the corresponding products and quarantine them. 2. Please complete the attached response letter and return it to ITP via E-Mail to the following address: Address: ITP GmbH Rckrufunterlagen Universittsstrae 136 44799 Bochum Germany Contact person: Dominik Frmke Tel.: +49 234 54621942 Fax: +49 234 54622444 E-Mail: fraemke@innotom.com 3. Upon receipt of the documents, we will contact you to retrieve and exchange the products. Information sharing: 4. Please inform all persons within your company belonging to the circle of users about this urgent safety