Elecsys Troponin I Immunoassay: Catalog# 05094810160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.

Company
Roche Diagnostics Operations, Inc.
Recall Initiated
April 7, 2021
Recall Number
Z-1596-2021
Quantity
726 kits
Firm Location
Indianapolis, IN

Reason for Recall

Recall reason is possible sporadic false negative conventional troponin results. Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/proBNP II STAT on the cobas e 601 and 602 modules. When the proBNP II analyte is measured immediately before the Troponin I analyte on the same module, a decreased recovery on Troponin I was observed.

Distribution

US Nationwide distribution in the states of AL, AZ, CA, CT, FL, GA, ID, IL, IN,KS, KY, MD, ME, MO, NC, NE, NJ, NM, NY, OH, OK, SC, SD, TN, TX, VA, WA, WY.

Lot / Code Info

UDI: 04015630939374

Root Cause

Device Design

Action Taken

On 04/07/2021, Roche issued Urgent Medical Device Correction via fax notifying customers that Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/ proBNP II STAT on the cobas e 601 and 602 modules. When the proBNP II analyte was measured immediately before the Troponin I analyte on the same module, a decreased recovery of Troponin I was observed.