RecallDepth

Speed Stitch Needle Cassette

Company
Arthrocare Corporation
Recall Initiated
August 17, 2020
Recall Number
Z-2999-2020
Quantity
6,366 needles
Firm Location
Austin, TX

Reason for Recall

Due to a supplier error where a portion of the raw material was inadvertently mixed. As a result, some needles may be made of commercially pure titanium as opposed to 304 stainless steel which does not meet the specification and could potentially break during use.

Distribution

Domestic Distribution: AK, AL, AR, AZ, CA, CO, CT,DB,FL,GA, HI, IA, ID, IL, IN, KS, KY, MA,MD,ME,MI,MN,MO,MS,MT, NC,ND, NE, NJ, NSW, NV,NY, OH,OK, OR, PA,SC, TN,TX, UT, VA, WA, WI, WV. International Distribution: AE, AU, CH, DE, FI, FR, GB, HK, IE, PA, PR, PT, RU, SG, and ZA

Lot / Code Info

Model: OM-8850 Lot Numbers: 2018434, 2018934, 2019558, 2019559, 2019929, 2019930, 2020931, 2022111, 2022641, 2030008, 2032212, 2033815, 2037604, 2036941, 2038701, 2039699, 2040500, 2042295, 2044957, 2047483, 2048575, 2051239 & 2032526

Root Cause

Mixed-up of materials/components

Action Taken

On August 17,2020 the firm notified all its consignees via letter of the field action. They provided the following instructions: Please inspect your inventory and locate any devices from the listed product and batch numbers, and quarantine them immediately. " If you are a distributor, you must notify your customers of the field action and ensure that these actions are carried out. If you have no product to return, please put an X in the appropriate location. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes. Complete the remainder of the form, sign and send to FieldActions@smith-nephew.com or fax to +1-901-566-7975.