CooperSurgical Dilator/Sound Set, Rx only, Sterile/EO, Made in South Korea, P/N MX21

Company
Coopersurgical, Inc.
Recall Initiated
April 9, 2020
Terminated
May 11, 2021
Recall Number
Z-2112-2020
Quantity
1740 units/174 boxes
Firm Location
Trumbull, CT

Reason for Recall

The seal of the sterile pouch may be compromised on one lot of the Milex Cervical Dilators.

Distribution

Distribution to North America including the following states: AZ, CA, CT, DC, DE, FL, GA, IA, IN, MA, MI, MN, MO, NC, NH, NY, OH, OK, PA, SC, TN, TX, UT, VA, WI

Lot / Code Info

Lot # 257589

Root Cause

Packaging process control

Action Taken

CooperSurgical sent an Urgent Medical Device Recall letter dated April 9, 2020 to all customers that received lot number 257589 of the Milex Cervical Dilator P/N MX21 which could have a compromised finished seal (opposite the chevron seal) of the sterile pouch. Customers were instructed to inspect each package for damage, including the seal area, prior to use (product is acceptable for use if it is visually confirmed that the pouchs seal is intact). Customers should discontinue use of and quarantine products with any packaging irregularities. The firm requested that the customers complete and return the attached Acknowledgement and Receipt Form, whether or not they had any of the affected product on hand. CooperSurgical stated that they will arrange for product returns and replacement.