NaviNetics Skull Anchor Key and Drill Kit, REF NN1215

This recall is currently active, issued April 4, 2024. It was issued by Navinetics Inc.

Company
Navinetics Inc
Recall Initiated
February 26, 2024
Posted
April 4, 2024
Recall Number
Z-1478-2024
Quantity
20 kits
Firm Location
Rochester, MN
Official Source
View on FDA website ↗

Reason for Recall

Defective Skull Anchor Key. After attachment of the single use Skull Anchor Key to the patient, the user was unable to attach the CT Localizer to perform stereotactic Imaging. Attachment of the localizer is typically accomplished by fully screwing in the four thumbscrews of the localizer into the threaded holes found on the top surface of the Skull Anchor Key.

Distribution

US Nationwide distribution in the states of FL, MN, CO.

Lot / Code Info

UDI/DI 10850023620276, Lot Number 331401724

Root Cause

Process control

Action Taken

NaviNetics issued an URGENT: NN1215 NAVINETICS SKULL ANCHOR KEY AND DRILL KIT RECALL notice to its consignees on 02/27/2024 via email. The notice explained the problem with the device, risk, and requested the return of the product.