RecallDepth

Impella catheters - Intravascular micro axial blood pumps that support a patient's circulatory system. (1) Product Code 005042 - Impella 2.5 (2) Product Code 005062 - Impella 5.0 (3) Product Code 005082- Impella LD (4) Product Code 004334 - Impella RP (5) Product Code 0046-0035 - Impella RP with SmartAssist (6) Product Code 1000323 - RP Flex with SmartAssist (7) Product Codes 0550-0008 and 1000100 - Impella 5.5 with SmartAssist (8) Product Code 0048-0032 - Impella CP (9) Product Codes 100080 and 0048-0045 - Impella CP with SmartAssist -- ***Added 2/20/24*** (10) 4413 and 004413 - Impella 2.5 Set (11) 5040 and 005040 - Impella 2.5 IMC Pump Set EU (12) 5060 and 005060 - Impella 5.0 IMC Pump Set EU (13) 5064 and 005064 - Impella 5.0 IMC Pump Set Canada (14) 1000115 - Impella CP Pump set, APAC (15) 1000211 - Impella 5.5 Smart Assist Set, JP (16) 1000302 - Impella CP with SmartAssist APAC (17) 1000361 - Impella 5.5 Set AU (18) 1000402 - Impella CP Smart Assist Set (19) 0046-0011 - Impella RP Pump Set, EU (20) 0046-0026 - Impella 5.0 Pump Set ROW (21) 0046-0037 - Impella 5.0 Pump Set APAC (22) 0046-0039 - Impella RP Set APAC (23) 004690-AU - Impella 5.0 Pump Set AU (24) 0048-0002 - Impella CP Pump Set, EU (25) 0048-0002-BR - Impella CP Pump Set BR (26) 0048-0004 - Impella CP Pump Set, Canada (27) 0048-0014 - Impella Smart Assist Set, EU (28) 0048-0024-JP - Impella CP Smart Assist Set, JP (29) 0048-0044 - Impella CP Smart Assist Set, Canada (30) 0048-0047- Impella CP Smart Assist Set APAC (31) 005048-JP - Impella 2.5 Pump Set, Japan (32) 005066-JP - Impella 5.0 Pump Set, Japan (33) 0550-0002 - Impella 5.5 with SmartAssist Set, EU (34) 0550-0004 - Impella 5.5 with SmartAssist Set, CA

Company
Abiomed, Inc.
Recall Initiated
December 27, 2023
Posted
February 12, 2024
Recall Number
Z-1074-2024
Quantity
92515
Firm Location
Danvers, MA

Reason for Recall

New warnings are being added to the Instructions for Use for all affected products to reduce the possibility of fibers being drawn into the Impella, which may result in low flow of the device.

Distribution

Worldwide distribution - US Nationwide and the countries of AU, CA, DE, FR, IN, MX, TW, and VI.

Lot / Code Info

UDI-DI (1) 00813502011081 (2) 00813502011180 (3) 00813502011227 (4) 00813502011029 (5) 00813502011869 (6) 00813502012811 (7) 00813502011531 and 00813502012828 (8) 00813502011388 (9) 00813502012279 and 00813502011876 All products are affected by IFU Update ***Added 2/20/24*** (10) 813502010947 (11) 4260113630136 (12) 4260113630174 (13) 813502011197 (14) 813502010473 (15) 813502012453 (16) 813502012767 (17) 813502012927 (18) 813502012873 (19) 4260113630273 (20) 813502011821 (21) 813502011937 (22) 813502011951 (23) N/A (24) 4260113630242 (25) 813502011777 (26) 813502011265 (27) 4260113630280 (28) 813502011609 (29) 813502012200 (30) 813502011944 (31) 813502010046 (32) 813502010053 (33) 813502011630 (34) 813502010466

Root Cause

Labeling Change Control

Action Taken

On December 27, 2023, the firm notified affected customers via URGENT MEDICAL DEVICE CORRECTION letters. Customers were informed of the IFU update, which included a New Warning: "To reduce the possibility of fibers being drawn into the Impella, customers should avoid exposing the inlet and cannula section of the Impella Heart Pumps to any surfaces or fluid baths where the device can come into contact with loose or floating fibers." The IFU udpate also includes updates to Potential Adverse Events and Warnings to add potential of cardiac or vascular injury. That part of the recall is classified and tracked under RES 93749. Customers should review the IFU updates and send their response form to the recalling firm. Please forward the recall notice to all impacted parties at your facility as well as any facilities to whom subject product has been forwarded. Post a copy of the notice in a visible area for awareness. Customers with questions may contact recallcoordinators@abiomed.com.

More recalls by Abiomed, Inc.

View all recalls by this company →