GIF-HQ190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE

This recall is currently active, issued November 29, 2023. It was issued by Aizu Olympus Co., Ltd. 3 Chome 1-1 Niiderakita Aizuwakamatsu Japan.

Company
Aizu Olympus Co., Ltd. 3 Chome 1-1 Niiderakita Aizuwakamatsu Japan
Recall Initiated
October 16, 2023
Posted
November 29, 2023
Recall Number
Z-0417-2024
Quantity
24 incorrectly repaired units
Official Source
View on FDA website ↗

Reason for Recall

Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.

Distribution

Domestic distribution to FL, KY, MA, NJ, NY, and PA

Lot / Code Info

UDI-DI 04953170305276 Serial Numbers 2270393, 2201027, 2201611, 2745407

Root Cause

Process control

Action Taken

On October 16, 2023, the firm notified affected customers through URGENT MEDICAL DEVICE REMOVAL letters. Customers were instructed to inspect their inventory and identify affected devices. Affected devices should be quarantined and not used. Olympus Customer Solutions will contact customers by phone and email to arrange the return and repair of the affected product. If you require additional information, please do not hesitate to contact Olympus at 647-999-3203 or Cynthia.Ow@Olympus.com.

More recalls by Aizu Olympus Co., Ltd. 3 Chome 1-1 Niiderakita Aizuwakamatsu Japan