Atrium Oasis Dry Suction Water Seal Chest Drain, Part Number: 3650-100, DRAIN, OASIS BRU W/AC, 3650 ATS BLOOD RECOVERY

This recall is currently active, issued December 2, 2023. It was issued by Atrium Medical Corporation.

Company
Atrium Medical Corporation
Recall Initiated
September 5, 2023
Posted
December 2, 2023
Recall Number
Z-0486-2024
Firm Location
Merrimack, NH
Official Source
View on FDA website ↗

Reason for Recall

Product was re-processed and re-sterilized by a third party using packaging configurations and a sterilization process which are not approved by the manufacture and lack validation.

Distribution

US: VA

Lot / Code Info

UDI-DI: 20650862113017. Lots: 461817, 467572, 467572, 468617, 468855, 467191, 467849, 469401, 479253.

Root Cause

Labeling mix-ups

Action Taken

The recall notification was issued via Fed-Ex and email on October 5, 2023. It requests (1) immediate quarantine and return of all affected product inventory on hand. (2) for any product further distributed, notify customers that have received affected product to remove the affected devices from use and return them. Questions may be addressed to your Getinge representative or call Getinge Customer Support at (603) 880-1433 (press option 1), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).