Abbott NT2000iX Radiofrequency (RF) generator, REF RFG-NT-2000, lesioning generator

Company
Abbott
Recall Initiated
September 28, 2023
Posted
November 2, 2023
Recall Number
Z-0209-2024
Quantity
34 devices
Firm Location
Plymouth, MN

Reason for Recall

Certain NT2000iX RF generators were serviced using a tool that was subsequently found to be out of tolerance and therefore may not have provided accurate results for potential capacitance failure.

Distribution

US, Taiwan, and Colombia

Lot / Code Info

UDI/DI 05415067022417, serial numbers: 16930509, 16983330, 17583407, 17634837, 17783928, 17783930, 17905961, 17928413, 17954521, 17991999, 18047384, 18493711, 18522937, 18675068, 18690920, 18709734, 18742997, 2475-12, 2504-12, 2547-13, 2885-14, 3003-14, 3278-15, 3450-15, 3550-15, 17942357, 18003310, 18102335, 18132017, 18132092, 18657521, 18666647, 19182119, 19197174

Root Cause

Process change control

Action Taken

Abbott issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 09/28/2023 via FedEx 2nd day delivery. The notice explained the problem, risk, and requested return of the device for inspection. Those devices which pass, will be returned. Should your device(s) fail the inspection, your local Abbott Representative can advise the consignee of alternate commercially available products.