Masimo LNCS TC-I SpO2, Reusable Ear Sensor, REF: 2503
- Company
- Masimo Corporation
- Recall Initiated
- April 21, 2023
- Posted
- June 21, 2023
- Recall Number
- Z-1986-2023
- Quantity
- 123,745 units
- Firm Location
- Irvine, CA
Reason for Recall
TC-I tip clip sensor may produce readings outside the accuracy specification.
Distribution
US: AK , AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY OUS: Argentina, Australia, Austria, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Gibraltar, Guernsey, Hong Kong, Hungary, Iceland, India, Israel, Italy, Japan, Jersey, Jordan, Kenya, Lebanon, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Palestinian Territory, Poland, Portugal, Puerto Rico, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam.
Lot / Code Info
UDI-DI: (01)00843997003230 Lot #: 13HBT, 15DAV, 17FAN, 17HAX, 17J12, 17MDF, 17N09, 18A28, 18AVB, 18CUV, 18FQZ, 18J33, 18K48,18KSS, 18MBC, 19ALF, 19CAY, 19D84, 19HMB, 19N1C, 20CRT, 20FMM, 20KWK, 21CRG, 21EMP, 21FFY,21FWV, 21G2K, 21GTL, 21HWG, 21KGD, 21MEF, 22AVS, 22BSC, 22DGZ, 22DUG, 22ESP, 22GDZ, 22J4J,22K64, 22NYC, 23AAE, E21DXT, K16HCA, K16JFC, K17BCR
Root Cause
Nonconforming Material/Component
Action Taken
On 5/17/2023, Masimo issued a Urgent: Medical Device Recall Notice via FedEx. Masimo asked consignees to take the following actions: 1. Contact Masimo Technical Services to request return of the affected part and lot number(s) for replacements and/ or if you have part numbers or lot numbers not included in this communication and have questions about those parts. 2. Masimo s Technical Services is available Monday through Friday 6:00am -5:00pm PST for live assistance or 24 hours a day to leave a message and someone will contact you. Local contact information can be found at http://service.masimo.com. 3. Complete Attachment 1, then email it to CustomerNotice@masimo.com, or Fax it to 1-949-297-7575 by MM DD, 2023. 4. Any adverse reaction or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program online, by regular mail or by fax.