RecallDepth

COULTER DxH Diluent, REF 628017; diluent used for counting and sizing blood cells

Company
Beckman Coulter, Inc.
Recall Initiated
May 31, 2023
Posted
July 12, 2023
Recall Number
Z-2138-2023
Quantity
335, 448 units
Firm Location
Miami, FL

Reason for Recall

Specific diluent lot numbers showed that conductivity, osmolality, and pH were out of manufacturing specifications.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Bolivia (Plurinational State of), Brazil, Canada, Chile, Colombia, Ecuador, Hong Kong, Indonesia, Kuwait, Mexico, Panama, Peru, Puerto Rico, Taiwan, Trinidad and Tobago, Uruguay, and Viet Nam.

Lot / Code Info

UDI DI 150995901HEMDILUENT5R; Container Number: 0001-0401: Lot numbers: Lot Numbers Beginning with 3541: 3541500, 3541510, 3541520, 3541550, 3541610, 3541620, 3541670, 3541690, 3541750, 3541770, 3541790, 3541810, 3541830, 3541860, 3541880, 3541900, 3541910, 3541920, 3541930, 3541940, 3541950, 3541990; Lot Numbers Beginning with 3542: 3542010, 3542030, 3542050, 3542070, 3542110, 3542150, 3542170, 3542180, 3542190, 3542230, 3542250, 3542260, 3542270, 3542310, 3542350, 3542370, 3542450, 3542490, 3542510, 3542530, 3542570, 3542590, 3542610, 3542630, 3542650, 3542670, 3542720, 3542740, 3542760, 3542780, 3542800, 3542820, 3542840, 3542850, 3542900, 3542920, 3542940, 3542960, 3542980; Lot Numbers Beginning with 3543: 3543000, 3543020, 3543040, 3543050, 3543080, 3543120, 3543140, 3543160, 3543170, 3543180, 3543200, 3543220, 3543240, 3543280, 3543300, 3543320, 3543340, 3543360, 3543370, 3543390, 3543410, 3543430, 3543460, 3543480, 3543500, 3543540, 3543560, 3543580, 3543610, 3543850, 3543870, 3543890, 3543910, 3543930, 3543950, 3543970, 3543990; Lot Numbers Beginning with 3544: 3544010, 3544030, 3544040, 3544050, 3544070, 3544090, 3544110, 3544130, 3544150, 3544160, 3544170, 3544190, 3544210, 3544230, 3544250

Root Cause

Under Investigation by firm

Action Taken

Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice on 05/31/2023 by email and mail. The notice explained the issue, its impact on the patient, and requested the following actions be taken: "Inspect the lot number set up on your instrument(s) and your current inventory with the corresponding container number (see Appendix A). Locate the container number on your diluent reagent 2D or Linear barcode label (see Appendix B). Contact Beckman Coulter Customer Technical Support Organization to replace the affected containers in your current inventory. Consult with your laboratory management to determine whether a retrospective review of results is necessary.