BALLARD* ACCESS Closed Suction System for Neonates/Pediatrics, Elbow Manifold.

This recall has been terminated (originally issued March 13, 2023).

Company
Avanos Medical, Inc.
Recall Initiated
February 3, 2023
Posted
March 13, 2023
Terminated
September 5, 2024
Recall Number
Z-1190-2023
Quantity
220 units
Firm Location
Alpharetta, GA
Official Source
View on FDA website ↗

Reason for Recall

Potential for product manifold to crack during use on patients, possibly causing leaks in the respiratory circuit and can lead to the patient receiving inadequate ventilation.

Distribution

US Nationwide distribution in the state of Oklahoma.

Lot / Code Info

All lots are affected. Stock Code (Lot Number): B50511 (20098277), B50611 (20098278), B50711 (20098279), B50811 (20097177, 20097177R1).

Root Cause

Component design/selection

Action Taken

Avanos hand-delivered the recall notification to the single consignee on 02/03/2023. The letter instructed the customer to discontinue use of and quarantine their stock of all lots of affected units and return them. Additionally, the customer was instructed to complete and return the acknowledgement form.