Finished Goods Box of 50 Inserts and Scanning Labels for Medivac Guardian 2L Catalog Number: FG 12010. Intended adjunct in the estimation of blood loss.

Company
Stryker Corporation
Recall Initiated
January 25, 2023
Posted
February 27, 2023
Recall Number
Z-1184-2023
Quantity
101 units
Firm Location
Portage, MI

Reason for Recall

Triton Canister Software, Insert & Scanning Label may cause the Triton Canister System to output inaccurate blood loss estimates can lead to confusion and could potentially result in either delayed recognition or prolongation of postpartum hemorrhage (PPH), or require more aggressive treatment for PPH.

Distribution

US Nationwide distribution.

Lot / Code Info

UDI-DI: 10859506006125 Lot Numbers: M2110004, M2201002, M2202001, M2203002, M2206001

Root Cause

Under Investigation by firm

Action Taken

Stryker issued Urgent Medical Device Correction letter to Risk Manager, OR Director, Materials Manager dated Jan 25, 2023. Letter states reason for recall, health risk and action to take: Steps: 1. Immediately check your internal inventory to locate any affected Finished Goods Inserts and Scanning labels (FG 12010 and FG 12022). Quarantine and discontinue use of all affected products. 2. A Stryker representative will contact your facility to assist updating the Triton Canister software. 3. Return the enclosed Business Reply Form (BRF) even if you don t have any affected product on hand, via email to instruments.recalls@stryker.com. 4. If the BRF indicates that recalled product is currently on hand, a shipping label will be provided to return recalled product. Upon receipt of the recalled product, Stryker will arrange replacements. 5. Maintain awareness of this communication internally until all required actions have been completed within your facility. Contact: Sylvia Page | Stryker Instruments Recall Coordinator 269-501-8375 |Instruments.recalls@stryker.com

More recalls by Stryker Corporation

View all recalls by this company →