cobas c 311; cobas c 501, 502,and COBAS INTEGRA 400 plus Analyzer/Module-RF interference claims for the sTfR assay Catalog Number: 20763454122

This recall has been terminated (originally issued August 28, 2019).

Company
Roche Diagnostics Operations, Inc.
Recall Initiated
August 28, 2019
Terminated
October 2, 2020
Recall Number
Z-0108-2020
Quantity
6,494 units
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

Tina-quant Soluble Transferrin Receptor (STFR) Assays Updated Claims for Rheumatoid Factors Interference

Distribution

Nationwide

Lot / Code Info

All serial numbers

Root Cause

Under Investigation by firm

Action Taken

Roche issued notification to consignees by Urgent Medical Device Correction (TP-00754) sent via UPS (signature required) on 8/28/19. The UMDC will also be available on diagnostics.roche.com. The letter identifies the problem, risk to health and action to take: Be aware of new interference claims that are listed in this UMDC and that the rhEPO Therapy Monitoring Statement will be removed from the Method Sheets. The UMDC will serve as intermediate labeling until the updated method sheets are available. Complete all sections of the enclosed fax form. Contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions if you have questions