Argon Medical Devices Plugs, Male/Female Luer Lock Plug, REF 041295000A, 50/box, Sterile, Rx. The firm name on the label is Argon Medical Devices, Inc., Athens, TX.

This recall has been terminated (originally issued October 16, 2018).

Company
Argon Medical Devices, Inc
Recall Initiated
October 16, 2018
Terminated
June 15, 2021
Recall Number
Z-0292-2020
Quantity
Total number in recall for all products in dist: 49,130 units
Firm Location
Athens, TX
Official Source
View on FDA website ↗

Reason for Recall

The seal formed on the edge of the pouch may be incomplete, which could result in a loss of sterility to the product.

Distribution

Distribution was nationwide. There was military/government distribution. Foreign distribution was made to Canada, Australia, the Netherlands, and Spain.

Lot / Code Info

Lot numbers 11228890, 11230669, and 11231874

Root Cause

Equipment maintenance

Action Taken

The recalling firm began issuing letters dated 10/16/2018 on 10/16/2018 via registered mail identifying the issue and the action to be taken by the customer.