Abbott ARCHITECT Creatine Kinase, IVD, REF 7D63-21.
This recall has been terminated (originally issued July 18, 2018).
- Company
- Abbott Laboratories, Inc
- Recall Initiated
- July 18, 2018
- Terminated
- August 13, 2020
- Recall Number
- Z-1075-2019
- Quantity
- 6,187 kits kits
- Firm Location
- Irving, TX
- Official Source
- View on FDA website ↗
Reason for Recall
The product has a stability issue which may lead to an error code indicating "Unable to calculate result, reaction check failure" and/or Quality Control results out of range low.
Distribution
Distribution was made nationwide, including Puerto Rico. There was also government and military distribution. Foreign distribution was made to Algeria, Argentina, Australia, Austria, Belarus, Belgium, Bermuda, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Gaza & Jericho, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Kazakhstan, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Mexico, Moldova, Morocco, Netherlands, Netherlands (Curacao), New Zealand, Norway, Oman, Pakistan, Paraguay, Philippines, Poland, Portugal, Qatar, Republic of Yemen, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turkey, UAE, Uganda, United Kingdom, Uruguay, Vietnam, and Zimbabwe.
Lot / Code Info
Lot number 87697UN18, exp. 22OCT18; UDI (01)00380740012878 (17)181022(10)87697UN18; Lot number 99632UN18, exp. 12NOV18, UDI (01)00380740012878 (17)181112(10)99632UN18; and Lot 24358UN18, exp. 05DEC18, UDI (01)00380740012878 (17)181205(10)24358UN18.
Root Cause
Material/Component Contamination
Action Taken
Recall letters were issued via FedEx overnight priority mail on 7/18/2018.