RecallDepth

HEMASHIELD Knitted Double Velour Cardiovascular Fabric

This recall has been terminated (originally issued March 20, 2018).

Company
Getinge Us Sales Llc
Recall Initiated
January 3, 2018
Posted
March 20, 2018
Terminated
July 3, 2018
Recall Number
Z-1172-2018
Quantity
1
Firm Location
Wayne, NJ
Official Source
View on FDA website ↗

Reason for Recall

A probable mislabeling of the product: a different-sized product is referenced on the product packaging label (i.e. 2.5 cm (width) x 7.6 cm (length)) while the actual product inside the box is 2.5 cm (width) x 15 cm (length).

Distribution

FL

Lot / Code Info

UDI code: 00384401017899 Catalog number: M002000195140 Lot number: 16M21 Serial number: 1092238229

Root Cause

Labeling mix-ups

Action Taken

On January 3, 2018 an URGENT MEDICAL DEVICE RECALL NOTICE was issued to the (1) affected customer. A return authorization was issued with the recall letter, which is intended to be returned to the recalling firm. Questions or concerns can be directed to 1-800-528-7486