Titan Touch Infr Zero Ang 22cm, Catalog Number ES28222400; inflatable penile Prosthesis

Company
Coloplast Manufacturing Us, Llc
Recall Initiated
March 1, 2023
Posted
April 12, 2023
Recall Number
Z-1368-2023
Quantity
8 units
Firm Location
Minneapolis, MN

Reason for Recall

A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.

Distribution

Worldwide distribution - US Nationwide and the countries of Turkey, Italy, Germany, Belgium, France, United Kingdom, Spain, Portugal.

Lot / Code Info

UDI/DI 05708932539142, Lot Numbers: 8849563, 8849585, 8904187

Root Cause

Process change control

Action Taken

The firm issued an URGENT MEDICAL DEVICE VOLUNTARY RECALL notice on 03/01/2023 by UPS next day air. The notice explained the issue and the hazard to health. The notice requests the removal and return of the affected product. Distributors are instructed to notify their customers. For further questions, please contact the Coloplast Customer Service team directly at (800) 258-3476, or by email urology@coloplast.com.