ET III Nano-HA fixture System ET NH Dental Implant Model # ET3R5010B -Intended for Dental Implant Catalog #: AET3R5010B

This recall is currently active, issued May 18, 2023. It was issued by Hiossen Inc. Keystone Industrial Port Complex.

Company
Hiossen Inc. Keystone Industrial Port Complex
Recall Initiated
April 6, 2023
Posted
May 18, 2023
Recall Number
Z-1604-2023
Quantity
287 units
Firm Location
Fairless Hills, PA
Official Source
View on FDA website ↗

Reason for Recall

Mis-labeled length of the dental implant on the Recall Implant's package. Dental implant size D5.0 x L8.5mm, labeled as D5.0 x L10mm

Distribution

US Nationwide Distribution

Lot / Code Info

UDI-DI: 00814913023717 Lot Number: H1E22J263

Root Cause

Labeling Change Control

Action Taken

Hoissen Inc. initial contact verbally through sales rep. on April 7, 2023 and by "URGENT: MEDICAL DEVICE RECALL" Letter via Mail & email on 04/19/2023. Letter states reason for recall, health risk and action to take: 1. Do not use the Recall Implants. 2. Quarantine (separate) the Recall Implants in a different location, away from normal stock room. 3. Immediately fill out the Acknowledgement and Receipt Form* (Fax: 267-795-7031 or Email: recall@hiossen.com) *Response is required regardless if the Recall Implants have already been return or implanted. 4. Returned the Recall Implant(s) as Unusable Returns to Hiossen QA** **Contact your sales representative for assistance. 5. Replacement devices are available for delivery. If you need any further information or support concerning this recall/issue, please reach HIOSSEN: Monday through Friday, 8:00 AM to 4:30 PM, Eastern Time. Tel: 267-795-7031 Email: recall@hiossen.com