HAMILTON-C6, REF: 160021

Company
Hamilton Medical Ag Via Crusch 8 Bonaduz Switzerland
Recall Initiated
March 14, 2023
Posted
April 25, 2023
Recall Number
Z-1429-2023
Quantity
103

Reason for Recall

Software error causes, safety ventilation, in which ventilation continues in the "safety ventilation" mode with audible/visible alarm - patient inputs are not monitored, if the following coincide 1) A mode change to an adaptive mode (ASV, APVcmv, APVsimv, INTELLiVENT-ASV, (S)CMV+, SIMV+), and 2) The controller and/or humidifier is connected to the ventilator and is operational.

Distribution

US Nationwide distribution in the states of CA, MA, ID, NC, GA, FL, NV, PA, MO, KS, SC, NY, NM, WY, VA.

Lot / Code Info

Software Version: 1.2.1, UDI-DI: 07630002808590

Root Cause

Software design

Action Taken

On 3/14/23, correction notices were emailed to health care facilities and distribution partners. Customers were told the following: 1) Firm will reach out to schedule the software update. 2) See Operation's Manual, Section 7.7, Safety Ventilation. The device must turn off to exit Safety Ventilation. 3) Complete and return the confirmation form to bret.everett@hamiltonmedical.com 4) If the malfunction occurs take the following actions: a) Monitor your patient. b) Provide an alternative device for ventilation. c) Have the ventilator serviced. If you have questions or concerns, contact Technical Support at 1-800-426-6331, option #2, or email Reno.TechSupport@hamiltonmedical.com