GEO Instrument Kit for EPS Plate, REF: 70120001, part of the GEO FirstFUSE 1st MTP Joint Arthrodesis Plating System

This recall is currently active, issued April 17, 2023. It was issued by Gramercy Extremity Orthopedics.

Company
Gramercy Extremity Orthopedics
Recall Initiated
April 4, 2023
Posted
April 17, 2023
Recall Number
Z-1395-2023
Quantity
24
Firm Location
Richardson, TX
Official Source
View on FDA website ↗

Reason for Recall

Instrument Kits for EPS plates may contain drill guides with incorrect drill depth range of 30mm - 60mm; not the correct drill depth range of 10mm - 40mm.

Distribution

US Nationwide distribution in the states of CT, OR, NJ, AZ, CT, OH, AZ.

Lot / Code Info

UDI-DI B471701200010, Serial Number(Expiration Date): 1022000493(10/13/27), 1022000623(10/13/27), 1022000683(10/13/27), 1022000243(10/13/27, 1022000313(10/13/27), 1022000913(10/13/27), 1022001213(10/13/27), 1022000403(10/13/27), 1022000383(10/13/27), 1022000883(10/13/27), 1022000353(10/13/27), 1022000663(10/13/27), 1022000733(10/13/27), 1022000793(10/13/27), 1022000983(10/13/27), 1022000993(10/13/27), 1022000373(10/13/27), 1022001193(10/13/27), 1022000523(10/13/27), 1022000283(10/13/27), 1022000223(10/13/27), 1022000363(10/13/27), 1022000413(10/13/27), 1022000453(10/13/27)

Root Cause

Mixed-up of materials/components

Action Taken

On 4/4/23, recall notices were mailed to customers and distributors who were asked to do the following: 1) Quarantine affected kits so they are ready for removal by the firm s sales reps or other contact. 2) If devices were transferred or further distributed, notify customers, or provide a customer list, so the recalling firm can notify these customers directly. 3) Complete and return the response form to msimpson@gramercyortho.com Customers with questions can contact the firm at 855-436-2278, 9AM 5PM, Monday - Friday.