Elekta Medical Linear Accelerator (ELMA) included in the below platforms: a. Elekta Infinity; b. Versa HD; c. Elekta Synergy

Company
Elekta, Inc.
Recall Initiated
May 5, 2023
Posted
May 30, 2023
Recall Number
Z-1665-2023
Quantity
46 units
Firm Location
Atlanta, GA

Reason for Recall

A component within a limited number of Elekta Linacs has been realized to be non-compliant to IEC 60601

Distribution

Worldwide distribution. US distribution to AR, IN, LA, OK, TN, VA, WA. International distribution to Belgium, Canada, Chile, China, France, Germany, India, Iraq, Italy, Jordan, Libyan Arab Jamahiriya, Mauritius, Mexico, Morocco, Poland, Russian Federation, Spain.

Lot / Code Info

Varex Part Number 95510354H-01 (Varex serial numbers 354-01H_A0010004PC to 354-01H_A0010269PC). a. Elekta Infinity: UDI-DI Number: 05060191071543, 05060191071512 (Serial Numbers: 156073, 152223, 153681); b. Versa HD: UDI-DI Number: 05060191071574, 05060191071529 (Serial Numbers: 156941, 156989, 156724, 156998); c. Elekta Synergy: UDI-DI Number: 05060191071550, 05060191071505 (Serial Number: 152483) Additional serial numbers distributed OUS: 109331, 109333, 109335, 109336, 109339, 109341, 109342, 151879, 152194, 153038, 153299, 153410, 153970, 156841, 156849, 156966, 156968, 156969, 156970, 156971, 156973, 156977, 156978, 156979, 156980, 156981, 156982, 156983, 156984, 156985, 156986, 156987, 156988, 156990, 156992, 156995, 156996, 156997.

Root Cause

Under Investigation by firm

Action Taken

Elekta notified customers, via email, on 05/05/2023. Customers were instructed that Elekta will contact each impacted customer to perform an onsite inspection of the power supply unit and replace as needed. Customers were asked to complete and return the Acknowledgement Form.