RecallDepth

Access-9 Large Bore Hemostasis Valves, Catalog Code: 580775001, IN8112/C, K05-00053H, K05-00109G, K05-00137K, K05-00410G, K05-00438G, K05-00502F, K05-00587D, K05-01052, K05-01706C, K05-01880K, K05-02130, K05-02595, K09-12599A, K10-05487P, K12-00264, MAP111/F, MAP112/B, MAP112/F, K08-YP085A, K08-YS017B, K08-YS055A, K09-YS029A, K09-YS040F, K12-YP002A, K12-YP030A, K12-YP060A, K12-YP061A, K12-YP063C, K12-YP072, K12-YS018A, K12-YS019A, K12-YS026A, K05-01140D, K05-01706C***, K05-02595, MAP112, K05-00564L, K12-YP002A,

Company
Merit Medical Systems, Inc
Recall Initiated
March 1, 2022
Recall Number
Z-0903-2022
Quantity
36445
Firm Location
Jordan, UT

Reason for Recall

A design change made to a hemostasis valve to improve manufacturability caused an internal gap between the rotator and the Y-body, where the guidewire can get caught, resulting in difficulty advancing the guidewire through the device. This defect may result in a delay of procedure.

Distribution

US: MN, PA, AZ, FL, IL, AK, NY, CA, TX, KY, WI, NC, NJ, VA, CO, IN, AR, MO, LA, OR, MT, MA, MI, NE, WV, KS, MS, HI, IA, NH, ME, AL, DC, DE, NM, OH, GA, MD, TN, UT, RI, SC, OK, WA, ID, ND, SD, NV, VT. OUS: Poland, Germany, France, United Kingdom, Belgium, Korea, Republic of, Austria, Canada, Spain, Luxembourg, Italy, Switzerland, Czech Republic, Netherlands, Japan, Honduras, Colombia, Greece, Ireland

Lot / Code Info

Catalog Code/Lots: 580775001/I2219771, I2278741, I2226790; IN8112/C/H2292357, H2287440, H2281404, H2298626, H2305047, H2305048, H2309356; K05-00053H/H2305217; K05-00109G/H2304656; K05-00137K/H2304653; K05-00410G/H2305216; K05-00438G/H2309130; K05-00502F/H2309122; K05-00587D/H2309211; K05-01052/H2304607; K05-01706C/H2309133; K05-01880K/H2306962; K05-02130/H2304639; K05-02595/H2304614; K09-12599A/H2295450; K10-05487P/H2279857; K12-00264/H2285040; MAP111/F/I2259670, I2279884, I2259657, I2289610; MAP112/B/P1908667; MAP112/F/I2259661, I2279915, I2279894, I2176725, I2215955, I2289611, I2279895, I2289613, I2289612; K08-YP085A/H2249051, H2302394; K08-YS017B/H2249052; K08-YS055A/H2275990; K09-YS029A/H2266470, H2302474; K09-YS040F/H2286601; K12-YP002A/H2269321, H2278279, H2283742, H2295661; K12-YP030A/H2263653; K12-YP060A/H2272013; K12-YP061A/H2249088; K12-YP063C/H2272382, H2295556; K12-YP072/H2266224; K12-YS018A/H2284956; K12-YS019A/H2302506; K12-YS026A/H2276014, H2302507; K05-01140D/H2304611; K05-01706C/ (H2304610, H2309133)***; K05-02595/H2304614; MAP112/I2279895; K05-00564L/H2304617; K12-YP002A/H2295661;

Root Cause

Component change control

Action Taken

On 03/01/22, recall notices were mailed to customers wo were informed to take the following actions: 1. Quarantine and discontinue use, and distribution of affected devices. 2. Ensure that applicable personnel within their organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, ensure the recall notice is immediately shared with them and note the quantity distributed on the Customer Response Form. 4. Fill out, scan, and email the completed Customer Response Form to Customer Service at response@merit.com. All affected product shipped to you must be accounted for on the Customer Response Form. 5. Immediately return all affected lots to the recalling firm. Customers with additional questions are encouraged to contact a sales representative or the firm s customer service at RESPONSE-EMEA@merit.com, 1-800-356-3748, Hours: 6 am to 6 pm MST, Mon-Fri. Recall notices were also sent to Sales representatives who were asked to contact their affected accounts and assist with product return and completion of the Customer Response Form. On 03/21/22, recall notices and response forms including an expanded scope were sent to customers informing them of additional distributed amounts of affected devices.

More recalls by Merit Medical Systems, Inc

View all recalls by this company →