YSIO X.Pree

Company
Siemens Medical Solutions Usa, Inc
Recall Initiated
March 7, 2022
Recall Number
Z-0730-2022
Quantity
53 units
Firm Location
Malvern, PA

Reason for Recall

For the automated multi-image-acquisition procedure Ortho x-ray collimation is set in a preparative stage for the entire examination area prior to the exam. During acquisition of each individual x-ray image, the x-ray collimator is automatically positioned in a way that the subsequent series of acquisition covers the defined field of view needed for each step. However, during the acquisition the collimation area displayed to the operator on the User Interface does not correctly represent the collimation area specified by the system. It indicates to the user an open collimator instead, e.g., the information displayed on the User Interface shows wider area of collimation than values preset prior to the examination. However, the collimation of the x-ray is performed correctly and always matches the examination area predefined by the user.

Distribution

US Nationwide, Worldwide distribution

Lot / Code Info

11107464

Root Cause

Software design

Action Taken

Siemens Medical Solutions USA, Inc. has initiated a Customer Advisory Notice to all affected customers. The letter inform customers that the firm will distribute a software correction via update instruction at no cost to the customers.