RecallDepth

TomoH(R) System, Model Number 1018284 Product Usage: The TomoTherapy treatment system is intended to be used as an integrated system for t he planning and precise delivery of radiation therapy, stereotactic radiotherapy, or stereotactic radiosurgery to tumors or other targeted tissues while minimizing the delivery of radiation to vital healthy tissues. The megavoltage x-ray radiation is delivered in a rotational, non-rotational, modulated (IMRT), or non- modulated (non-IMRT/3 dimensional conformal) format in accordance with the physicians prescribed and approved plan.

This recall has been terminated (originally issued November 17, 2017).

Company
Accuray Incorporated
Recall Initiated
November 17, 2017
Terminated
July 25, 2021
Recall Number
Z-0288-2018
Quantity
29 units
Firm Location
Madison, WI
Official Source
View on FDA website ↗

Reason for Recall

Uncontrolled couch Z-axis movement (descent)

Distribution

Worldwide Distribution

Lot / Code Info

Serial Numbers: 110040 110083 110163 110231 110237 110289 110302 110369 110418 110461 110462 110474 110491 110494 110503 110519 110525 110533 110536 110537 110552 110554 110556 110562 110565 110570 110573 110576 110577

Root Cause

Software design

Action Taken

Accuray sent an Urgent Field Safety Notice letter dated November 8, 2017, to affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated the following: "Accuray(R) will correct all affected High-Performance Patient Couches. An Accuray representative will contact you to arrange a time to update your system. The update includes the installation of TomoTherapy(R)' H(TM) Series software version 2.1.2 (Hi Art(R) vS.1.2) on your system, as well as the replacement of cables within the couch subsystem." For questions contact Accuray Customer Support by phoning the regional support lines.