RecallDepth

St. Jude Medical. Agilis NxT Steerable Introducer: B1-Directional, Medium Curl 22.4 mm, 71 cm, 6.5 F ID; Dilator: 94 cm, 8.5 F; Super Stiff, Guidewire: 180 cm, 0.032". Reprocessed device for single use only, sterile. Product Usage: The reprocessed steerable introducer sheaths are indicated when introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.

This recall has been terminated (originally issued December 29, 2017).

Company
Sterilmed Inc
Recall Initiated
June 2, 2017
Posted
December 29, 2017
Terminated
July 29, 2019
Recall Number
Z-0213-2018
Quantity
112 devices
Firm Location
Minneapolis, MN
Official Source
View on FDA website ↗

Reason for Recall

Inadequate seal due to lack of adhesive glue

Distribution

US Nationwide distribution in the states of AR, CO, FL, MI, OH, TX, and WA.

Lot / Code Info

Model No. STJ408309, STJ408310, STJ408324. Exp. Date of 05/31/2018 or PRIOR. All Lots, known lots: 7344024, 7343120, 7343643, 7343797, 7348305, 7355750, 7348712, 7290257, 7354522, 7354523, 7350001, 7324499, 7343501, 7351203, 7255673, 7273483, 7343603, 7355439, 7355745, 7355754, 7355752, 7282741, 7315539, 7320863, 7343766, 7351038, 7330645, 7345143, 7323722, 7330603, 7342081, 7351724, 7352687, 7355130, 7355131, 7355137, 7377760, 7355135, 7375662, 7404300, 7404301, 7404316, 7404306, 7343395, 7355463, 7355714, 7355622, 7330666, 7355572, 7327999, 7355677, 7308337, 7343686, 7355680, 7251903, 7259400, 7322815, 7342069, 7355574, 7355676, 7355712, 7346336, 7355666, 7355661, 7355618, 7355571, 7354437, 7350573, 7331054, 7300290, 7350792, 7355679, 7355709, 7355713, 7355681, 7355675, 7355573, 7345129, 7343838, 7355678, 7355575, 7355620, 7355711, 7346820, 7355662, 7404299, 7404282, 7404294, 7404290, 7404310, 7404311, 7404318, 7404297, 7404287, 7342672, 7357129, 7241712, 7350942, 7272027, 7316930, 7345592, 7348169, 7388731, 7372138, 7386929, 7387147

Root Cause

Process control

Action Taken

Sterilmed sent an Urgent Medical Device Recall letter dated June 12, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine inventory to determine if affected products are on hand, complete and return the Business Reply Form, share information with appropriate staff in facility, attach copy of this notification to quarantined product, and return affected product to Sterilmed. For questions contact Sterilmed's Customer Care Team at 888-541-0078.