ARCHITECT c4000 part number 02P24; ARCHITECT c8000 part number 01G06; ARCHITECT c16000 part number 03L77 Clinical Chemistry systems.

This recall has been completed (originally issued September 29, 2017).

Company
Abbott Laboratories, Inc
Recall Initiated
September 29, 2017
Recall Number
Z-0165-2018
Quantity
34 units
Firm Location
Irving, TX
Official Source
View on FDA website ↗

Reason for Recall

The Peristaltic Head tubing placed on the ARCHITECT c4000, c8000, and c16000 has the possibility to leak due to a manufacturing issue associated with the tubing connector.

Distribution

Worldwide distribution to US, Canada, France, and Germany.

Lot / Code Info

c400073, c400103, c400498, c400600, c400890, c400934, c400935, c401018, c401055, c401056, c401182, c401505, c401520, c401686, c401887, c460196, c460200, c460943, c461550, c803669, c803753, c1600144, c1600249, c1600251, c1600641, c1600720, c1600819, c1600859, c1600976, c1601350, c1601351

Root Cause

Component design/selection

Action Taken

All worldwide ARCHITECT Clinical Chemistry System customers with impacted instruments will receive the Product Correction Letter dated, 9/29/17. A customer reply is requested from all affected consignees.