DIASTAT(R) ANA (anti-Nuclear antibody) EURO DIAGNOSTICA, The DIASTAT(R) anti-nuclear antibody (ANA) test is a qualitative enzyme-linked immunosorbent assay (ELISA) for the detection of ANAs in human serum or EDTA plasma. It detects ANAs against Sm, Sm/RNP, Ro (SS-A), La (SS-B), Scl-70, Jo-1, dsDNA, histone and centromere antigens.

This recall has been terminated (originally issued May 22, 2017).

Company
Euro Diagnostica Ab
Recall Initiated
May 22, 2017
Terminated
March 31, 2021
Recall Number
Z-2627-2017
Quantity
501 units
Firm Location
Malmo Sweden
Official Source
View on FDA website ↗

Reason for Recall

FANA200 kit lot TS 3577 was re-worked due to shortened shelf-life. The alert sticker on the kit box of the lot informed customers that the shelf-life had been shortened and referenced a letter sent to customers for more information, where the new expiration date was stated. The original kit box label remained on the kit box stating the original expiration date, the packaging did not display the correct expiration date.

Distribution

MN, TX, NE, PA, Spain, Egypt, Pakistan, Nepal, Canada

Lot / Code Info

TS 3577

Root Cause

Other

Action Taken

The firm initiated the problem by email on 05/22/2017. The consignee was directed to destroy the product.

More recalls by Euro Diagnostica Ab