Olympus Leak Test Adapter used with Acu-sInQ Complete Endoscope Cleaning Aid Systems-intended to assist manual flexible endoscope cleaning Part number: 0600818OKN

Company
Steris Corporation
Recall Initiated
October 7, 2020
Terminated
July 14, 2021
Recall Number
Z-0458-2021
Quantity
293 units:248 units (USA); 45 units OUS
Firm Location
Mentor, OH

Reason for Recall

Potential for cleaning solution intrusion into endoscope

Distribution

Nationwide Foreign: Canada, Italy, Kuwait, United Kingdom, United Arab Emirates

Lot / Code Info

Distributed from May 23, 2017 to June 30, 2020

Root Cause

Device Design

Action Taken

Steris issued Urgent Medical Device Recall issued letter on 10/7/20 stating reason for recall, health risk and action to take: Please immediately inspect your facility for affected Leak Test Adapters that connect to Olympus endoscopes. The affected Leak Test Adapters have a white plastic valve connector. If you have this Leak Test Adapter, please disconnect from your Acu-sInQ Complete, destroy, and replace with the new Leak Test Adapter included with this Customer Notification Letter. 2. Please complete the Medical Device Recall Response Form included with this Customer Notification Letter. Your STERIS Sales Representative can assist you should you have any questions while completing the form, or when replacing the Leak Test Adapter. 3. Return the completed Medical Device Recall Response Form to STERIS via one of the following three methods: a. Visit https://www.steris.com/medical-device-recall-response-form and complete the form electronically; b. Scan and email the completed form to Regulatory_Compliance@STERIS.com; or c. Fax the completed form to (440) 392-8963. We apologize for any inconvenience this matter may cause, and as always, STERIS is dedicated to supporting our products and valued Customers. If you have questions regarding this matter, please contact Katilin Hawranko, Product Manager at (440) 392-7431, STERIS Customer Service at 1-800-548-4873, or your local STERIS Representative.

More recalls by Steris Corporation

View all recalls by this company →