RecallDepth

Dual Mobility Vivacit-E Bearing, Model Number 110031009

This recall has been terminated (originally issued August 14, 2020).

Company
Zimmer Biomet, Inc.
Recall Initiated
August 14, 2020
Terminated
June 9, 2022
Recall Number
Z-0057-2021
Quantity
31
Firm Location
Warsaw, IN
Official Source
View on FDA website ↗

Reason for Recall

The outer package is labeled as a Size C 38 mm, however, the implant inside the package is a Size E 42 mm.

Distribution

Products were distributed to the following US states: FL, ID, IL, IN, KY, MA, MD, MI, MN, NC, NJ, OH, OR, PA, SC, TX, UT, VA, WI, and WV.

Lot / Code Info

Lot Number 64755636

Root Cause

Under Investigation by firm

Action Taken

Urgent Medical Device Recall notification letters dated 8/14/20 were sent to customers. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form will be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. If the product has been implanted, Zimmer Biomet recommends assessing the patient with imaging and monitoring the patient for the potential health risks. 3. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred

More recalls by Zimmer Biomet, Inc.

View all recalls by this company →