Siemens Dimension Vista Akaline Phosphatase (ALPI) Flex reagent cartridge - In vitro diagnostic test for the quantitative measurement of alkaline phosphatase in human serum and plasma Catalog Number: K2115, SMN# 10642444

Company
Siemens Healthcare Diagnostics, Inc.
Recall Initiated
August 25, 2020
Posted
October 20, 2020
Terminated
May 4, 2023
Recall Number
Z-0134-2021
Quantity
US:5970 units ; OUS: 1446 units
Firm Location
Newark, DE

Reason for Recall

Low outlier resultsgenerated with Alkaline Phosphatase (ALPI) Flex reagent cartridge.The maximum difference observed with a patient sample was approximately -27% at a concentration of 129 U/L (2.15 ukat/L). The low outlier results are generated from the last five (5) tests of the well

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Australia, Austria, Belgium, Bahamas, Czech Republic, Denmark, France, Germany, Italy, Japan, Korea, Netherlands, Norway, New Zealand, Portugal, Slovakia, Slovenia, Spain, Switzerland.

Lot / Code Info

Lot Number/UDI/Exp. Date:  1) 19247AB  UDI: 0084276803903419247AB20090310642444840  Exp.2020-09-03, (2) 19282BB  UDI: 0084276803903419282BB20100810642444840  Exp. 2020-10-08, (3) 19330BD   UDI:0084276803903419330BD20112510642444840  Exp. 2020-11-25.  Expanded Recall: 3/10/21 20337AB, 20311BA, 20275AA, 20218AA, 20203BA, 20170AA, 20125BA, 20078AB.

Root Cause

Under Investigation by firm

Action Taken

Siemens issued Urgent Medical Device Recall (UMDC # VC-20-02.A.US via Fed'X) and Urgent Field Safety Notice (UFSN # VC-20-02.A.OUS) on August 25, 2020 to all Dimension Vista customers who received the Dimension Vista ALPI flex reagent lots 19247AB, 19282BB and 19330BD to notify them of the issue, health risk and actions to take. Customers were instructed to: Discontinue use of and discard the Dimension Vista ALPI lots 19247AB, 19282BB, 19330BD. Review their inventory of these lots to determine their laboratorys replacement needs. Complete and return the attached form to this letter to request your no-charge replacement product(s). Complete and return the Field Correction Effectiveness Check Form attached to the letter within 30 days. If they received any complaints of illness or adverse events associated with the product listed in the letter, to immediately contact their local Siemens Remote Services Center or their local Siemens Technical Support Representative. Expanded Recall 3/12/21: Urgent Medical Device Correction (UMDC # VC-20-02.B.US) and Urgent Field Safety Notice (UFSN # VC-20-02.B.OUS) issued on March 12, 2021 to all Dimension Vista customers who received the Dimension Vista ALPI flex reagent lots 20337AB, 20311BA, 20275AA, 20218AA, 20203BA, 20170AA, 20125BA and 20078AB to notify them of the issue, and actions to be taken.