This is an in vitro diagnostic product intended for use as a calibration serum in clinical chemistry assays. Randox calibration sera contain 42 analytes and are based on lyophilized human serum. The concentrations and activities are suitable for calibration of clinical chemistry assays both manually and on a wide range of automatic analyzers. Constituent concentrations are available at 2 levels.

This recall has been terminated (originally issued July 12, 2019).

Company
Randox Laboratories Ltd. 55 Diamond Road Crumlin (North) Ireland
Recall Initiated
July 12, 2019
Terminated
June 17, 2020
Recall Number
Z-2490-2019
Quantity
307 kits
Official Source
View on FDA website ↗

Reason for Recall

Randox Laboratories realigned RX Analyser Series calibration targets by 10% for AST and 9% for ALT in line with the Mean of All Instrument method target for the Randox Calibration Serum Level 3 (for calibrator lot 961UE onwards).

Distribution

Nationwide Foreign: Antigua, Aruba, Bahamas, Granada, Tobago,

Lot / Code Info

Lot Numbers: 961UE onwards GTN: 05055273200966

Root Cause

Under Investigation by firm

Action Taken

Randox Laboratories Ltd. issued Urgent Medical Device Correction letter on 7/12/19 advising users of the problem, health risk and action to take: Replace all Quality Control value sheets with the revised sheets available on www.randox.com . Discuss the contents of this notice with your Medical Director. Complete and return the response form 12187-QA to technical.services@randox.com within five working days.