Elekta Unity, Image-Guided Radiation Therapy System Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan.

Company
Elekta, Inc.
Recall Initiated
June 24, 2019
Posted
July 16, 2019
Terminated
April 1, 2021
Recall Number
Z-1980-2019
Quantity
12 units
Firm Location
Atlanta, GA

Reason for Recall

The QA software solution to perform the MR to MV alignment check, does not display the stored MR to MV offset values. The user is unable to independently inspect the values during their QA.

Distribution

US Distribution to states of: TX & WI, and Internationally to: Canada, Denmark, Germany, Italy, Netherlands, Sweden, and the UK.

Lot / Code Info

UDI/GTIN: 5060081071321 Installed product: UNITY SYSTEM/13553-006/600003 UNITY SYSTEM/10817-011/600010 UNITY SYSTEM/12473-020/600020 UNITY SYSTEM/12274-017/600012 UNITY SYSTEM/10562-011/600016 UNITY SYSTEM/10719-T03001/600013 UNITY SYSTEM/30004070-001/600023 UNITY SYSTEM/10420-AVL-U/600008 UNITY SYSTEM/11014-UMCU-U/600007 UNITY SYSTEM/10157-006/600014 UNITY SYSTEM/11611-40/600011 UNITY SYSTEM/11489-21/600009

Root Cause

Software design

Action Taken

The firm, Elekta, disseminated their "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" by email on 06/24/2019 to its customers. The notices identified the problem, provided recommended actions, and requested the notice be placed in a place accessible to all users. Customers were also instructed to complete and return the Acknowledgement Form. A software update will be provided once available. If you have any questions, contact the Post Market Surveillance Specialist by email: Elekta_Global_PMS@elekta.com.